| Model Number |
LF1520 |
| Medical Device Problem Codes |
Energy Output To Patient Tissue Incorrect (1209); Sticking (1597)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident device has been received and is under evaluation.When the device evaluation is complete a follow-up report will be submitted.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that during procedure, the device did not seal tissue properly from the beginning of use.The device was activated but the thermal spread did not seem to be enough.Tissue would also stick to the jaws, and the jaws were difficult to open.There was no patient injury, and a new device was used to continue the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
One used lf1520 ligasure device was received, and evaluation of the returned sample could not confirm the reported issue.Visual inspection found no defects.The device was activated multiple times on porcine kidney samples, pressing the button in various locations on a range of vessel sizes to detect any activation issues.All seal cycles were completed satisfactorily.A full thermal effect was visually verified without tissue sticking.The investigation found the device to function normally and within specifications.There are factors during use that can affect seal quality, and the instructions for use included with this product lists measures to ensure sealing effectiveness.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|