• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ESSENCE; COMPRESSOR, AIR, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ESSENCE; COMPRESSOR, AIR, PORTABLE Back to Search Results
Model Number 1112266
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/29/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this device stopped working.On return of the device for evaluation it was observed that the strain relief on the mains cord was not sitting in its recess, and the insulation on the mains cord was damaged to the extent that bare copper wires were exposed.Based on this, the patient may have been exposed to the risk of electrocution.The ifu for this device contains the warning 'the power supply cord cannot be replaced by the user.In case the power supply cord becomes damaged, contact philips respironics customer service for replacement (there is no service option)'.No patient harm or death has been reported and the device is not life sustaining/supporting.' (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ESSENCE
Common Device Name
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
MDR Report Key6763486
Report Number9681154-2017-00019
Device Sequence Number821922
Product Code CAF
UDI-Device Identifier00383730001951
UDI-Public00383730001951
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K042655
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1112266
Device Catalogue Number1112261
Device Lot Number151021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2017
Initial Report FDA Received Date08/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-