• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problems Device Operates Differently Than Expected (2913); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/21/2017
Event Type  malfunction  
Event Description
The recipient reportedly experienced intermittent lock and increased power consumption.External equipment was exchanged and programming adjustments were made, however, the issues did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
The external visual inspection revealed the electrode was severed near the fantail and near the array prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was obtained only at certain spacing.The electrode condition prevented several of the electrical tests from being performed.The device passed some of the electrical tests from being performed.The device failed the residual gas analysis test.This device had moisture that exceeded the residual gas analysis test limit.Leak testing did not reveal any leaks.Capas were implemented.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
pamela campo
28515 westinghouse place
valencia, CA 91355
6613627624
MDR Report Key6768327
MDR Text Key81857829
Report Number3006556115-2017-00412
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Audiologist
Type of Report Initial,Followup
Report Date 07/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/30/2006
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2017
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age20 YR
-
-