| Model Number |
680R |
| Medical Device Problem Codes |
Gradient Increase (1270); Inadequacy of Device Shape and/or Size (1583); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Mitral Regurgitation (1964); Mitral Valve Stenosis (1965); No Information (3190)
|
| Date of Event |
06/21/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the return of the product, no definitive conclusion can be made regarding the clinical observation.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that 1 year 2 months post implant of this mitral annuloplasty ring, the device was explanted and replaced with a bioprosthetic valve for unknown reasons.No other adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received indicated that the device was explanted due to symptomatic mitral stenosis, residual regurgitation, and elevated gradients.The physician stated that the ring was probably too small for the patient, and this, combined with underlying subvalvular fibrosis, led to the stenosis.No additional adverse patient effects were reported. added patient weight; added patient device codes.If information is provided in the future, a supplemental report will be issued.
|
| |
|