| Medical Device Problem Code |
Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a button/keypad (tactile changes/unresponsive) issue.It was reported that the down arrow was under-responsive.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue has the ability to result in inadvertent or incorrect insulin delivery or the inability to use the pump.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission: 10/02/2017.Device evaluation: the device has been returned and evaluated by product analysis on 09/09/2017 with the following findings: during investigation, the down arrow button was under responsive.The keypad was removed to find the down arrow button contact was cracked.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|