| Model Number |
48850000 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Injury (2348); No Code Available (3191)
|
| Date of Event |
07/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; revision l3-l5 after rods slipped out of screws.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed via visual inspection and x-ray images.No manufacturing issues were identified.A root cause of the event is overloading on one side of the blocker due to the blocker not being optimally seated on the rod.
|
| |
|
Event or Problem Description
|
|
It was reported that; revision l3-l5 after rods slipped out of screws.
|
| |
|