| Model Number |
840 |
| Medical Device Problem Code |
Device Inoperable (1663)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that an 840 ventilator became inoperative.At this time, it is unknown if there was patient involvement.A service engineer (se) inspected the device and replaced the gui (graphical user interface) printed circuit board (pcb) and updated the backlight inverter pcbs to address the issue.The unit passed all testing and operated within the manufacturing specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
A graphic user interface (gui) printed circuit board (pcb) and backlight inverters pcb were returned to covidien/medtronic¿s product analysis.A visual inspection of the returned component was performed, no notable conditions were found.The returned components were installed into a test ventilator for analysis; no errors were recorded in the diagnostic logs.An investigation was performed and the product analysis technician reported that no fault was found.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The ventilator was not in use on a patient at the time of the event.
|
| |
|