| Model Number |
B3847 |
| Medical Device Problem Code |
Knotted (1340)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/23/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ra has received the incident device and the product has been assigned to engineering for evaluation.A follow-up report will be sent upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: removal of stent.Incident description: stent inserted on (b)(6) 2017, removed (b)(6) 2017.Proximal end of stent "knotted" in renal pelvis.See ureter breached diathermy to extract stent.Further information from urologist.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Upon visual inspection, engineering confirmed the complainant's experience of knotting at the proximal end of the stent body.The wire was exposed in the location of the knot.The root cause of the reported event could not be determined.The formation of knots is rare and publications from medical literature have found that excessive length and presence of additional coils may contribute to knot formation in multi-length ureteral stents.As such, the instructions for use (ifu) includes the following warning, "formation of knots in multi-length ureteral stents may occur.This may result in injury to the ureter during removal and/or the need for additional surgical intervention.The presence of a knot should be considered if significant resistance is encountered during attempts at removal." applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.
|
| |
|
Search Alerts/Recalls
|