| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a power (damage) issue.It was alleged that the battery compartment was cracked with no power.There was no indication that the product caused or contributed to an adverse event.This issue is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
|
| |
|
Search Alerts/Recalls
|