|
A flexor ansel guiding sheath was used in the treatment of a chronic total occlusion (cto) lesion in the superficial femoral artery (sfa ).The procedure was performed by contralateral approach from the groin.There were calcifications noted in the puncture site (groin).After a 4fr.Short sheath, details unknown, was inserted into the artery, the physician attempted to advance the complaint device into the patient's body over a wire guide.It would not enter the artery and as a result, the tip became deformed, resulting in fraying.Another manufacturer's sheath was used instead, without problem.No adverse effects on the patient were reported do to this occurrence.
|
|
Investigation - evaluation: a review of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, quality control, and visual inspection of the returned device was conducted during the investigation.The flexor ansel guiding sheath was returned in a used condition and examined.The sheath was returned with the 0.038" dilator inserted into the proximal fitting of the sheath.The distal tip of the sheath appeared to be smooth tapered with smooth transition between sheath and dilator.The curve of the sheath was also within tolerance.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that would contribute to this failure mode.There was one other reported complaint for this lot number.Based on the available information, examination of the returned product, and results of the investigation, it is likely that the calcification present at the puncture site might have caused difficulties in getting the sheath into the patient's body.Per the quality engineering risk assessment, no further action is required.Appropriate personnel have been notified and monitoring will continue to be performed for similar complaints.
|