• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE COOK-COPE TYPE LOCKING LOOP BILIARY DRAINAGE CATHETER; GCA CATHETER, BILIARY, DIAGNOSTIC CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE COOK-COPE TYPE LOCKING LOOP BILIARY DRAINAGE CATHETER; GCA CATHETER, BILIARY, DIAGNOSTIC CATHETER, NEPHROSTOMY Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Code Available (3191)
Date of Event 11/01/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event or Problem Description
Report received via medsun report (b)(4) : "while removing a nephrostomy tube during a transplanted kidney nephrectomy after rejection last fall, the nephrostomy tube was cut as part of the procedure and the main part of the tubing was removed on the kidney side.A 4cm piece of the tubing that was in the bladder was not removed and thus retained in the patient.This was discovered during recent hospitalization on ct scan and it was removed yesterday." original intended procedure: left-sided transplant nephrectomy with nephrostomy tube removal." instructions for use that accompany each device provide instructions for the removal and exchange of the catheters.
 
Additional Manufacturer Narrative
Additional information: the device was requested; however, it is not being released per the facility's risk management department at this time.Investigation - evaluation: a review of complaint history, device history record, documentation, instructions for use (ifu), and quality control data was conducted during the investigation.The complaint device was not returned; therefore, device failure analysis and physical examination of the device used in this case could not be performed.The manufacturing documents in place at the time of manufacture were reviewed and it was found that the device aspect in question was visually inspected by quality control and no notable gaps in production or processing controls were identified.The device history record for the complaint lot number was reviewed.No related non-conformances were identified.A search of our complaint records indicates that this is the only complaint on the lot number at the time of investigation.There is no evidence to suggest that this device was made out of specification.Based on the information provided by the customer, the nephrostomy tubing was cut during the surgical removal of the kidney and inadvertently left in the bladder.Therefore, the root cause was determined to be user technique.We will notify the appropriate personnel and continue to monitor for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE COOK-COPE TYPE LOCKING LOOP BILIARY DRAINAGE CATHETER
Common Device Name
GCA CATHETER, BILIARY, DIAGNOSTIC CATHETER, NEPHROSTOMY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6777555
Report Number1820334-2017-02242
Device Sequence Number220257
Product Code GCA
UDI-Device Identifier00827002092051
UDI-Public(01)00827002092051(17)190527(10)7002343
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberULT16.0-38-40-P-32S-BCL-HC
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/26/2017
Supplement Date Received by Manufacturer11/28/2017
Initial Report FDA Received Date08/09/2017
Supplement Report FDA Received Date12/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
-
-