NEWPORT MEDICAL INSTRUMENTS, INC HT70 PLUS VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number HT70
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that an ht70 plus ventilator displayed backup battery failure and had a coin battery problem after turning on the device.The ventilator was not in use on a patient at the time of the reported event.A service engineer (se) inspected the device and replaced the control board and coin battery to resolve the issue.The unit passed all testing and operated within the manufacturing specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HT70 PLUS VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
NEWPORT MEDICAL INSTRUMENTS, INC
1620 sunflower ave
costa mesa CA 92626
MDR Report Key6777845
Report Number2023050-2017-05641
Device Sequence Number1539575
Product Code CBK
UDI-Device Identifier10884521542181
UDI-Public10884521542181
Combination Product (Y/N)N
Initial Reporter CountryCS
PMA/510(K) Number
K111146
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHT70
Device Catalogue NumberHT70
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2017
Initial Report FDA Received Date08/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-