| Model Number |
G23382 |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The physician reported during stent placement using a 6 fr universa soft ureteral stent set, when placing the stent with the aid of the monofilament string, the string kept stretching and finally the string broke.A 7 fr universa soft ureteral stent set was then used and the same issue occurred; the monofilament string stretched and finally broke.As reported, no additional procedures were required as a result of these issues.No section of the device was left inside the patient's body.There were no adverse effects to the patient due to these issues.Additional patient and event detail has been requested but not been received at the time of this report.As reported, there were no adverse effects to the patient due to this occurrence.
|
| |
|
Event or Problem Description
|
|
This event has been reported in two related manufacturer reports to reflect there were two device malfunctions reported to occur during the same procedure.The first report mfr.Report # 1820334-2017-02293 captures the first reported device issue.The second device issue has been captured in mfr.Report # 1820334-2017-02294.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation ¿ evaluation: as anticipated, the universa soft ureteral stent set was not returned for evaluation.No photos have been provided.Without the complaint device, a physical investigation was not able to be completed.Based on the available information a definitive root cause could not be established at this time.There is no indication that a design or process related failure mode contributed to this event.A review of production and quality documentation did not identify any specific issues with current manufacturing or quality controls that may have contributed to this incident.The device history record was reviewed and no non-conformances were noted.A review of complaint history for this product/lot number combination revealed this complaint is the only reported complaint associated with this complaint lot number.Per the quality engineering risk assessment, no further action is required.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.
|
| |
|