| Model Number |
2319 |
| Medical Device Problem Codes |
Device Alarm System (1012); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/31/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned sensor was evaluated.During inspection, the sensor passed software testing and visual analysis.The sensor failed continuity testing due to broken green and white conductors at the detector solder joint assembly.During testing, the sensor was able to generate audible and visual alarms, and an error message indicating "probe is off the patient." the reported issue was not confirmed.
|
| |
|
Event or Problem Description
|
|
The customer reported the following issue: "if the probe goes bad (we have had a few), there is no alarm, the patient is then hooked up with no sats on the monitor." no consequences or impact to patient were reported.
|
| |
|
Search Alerts/Recalls
|