ABBOTT MOLECULAR, INC. ABBOTT REALTIME HIV-1 AMP REAGENT KIT; TEST, HIV DETECTION Back to Search Results
Model Number 6L18-90
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/22/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint ticket has been reviewed and has been determined to be the same issue as previously confirmed issue (b)(4) capa.Complaint investigation included customer data review, product evaluation and complaint history review; summary included below: customer data review: customer complaints were confirmed when am technical product support (tps) identified multiple error codes (error code 4440 - "positive control concentration is out of range", error code 4441 - "internal control cycle number is too low" and error code 4457 - "internal control failed") when reviewing customer run data and internal testing of retained file samples of lot 476139.(b)(4) capa has been opened to investigate this issue.Product evaluation: product evaluation with file samples of the abbott realtime hiv-1 amp rgt kit-row (list 02g31-010) lot 476139 and abbott realtime hiv-1 control kit-row (list 02g31-80) lot 473517 used by the original complainant identified the issue relating to control concentrations out of range with early ic ct response for these replicates that could be related to the reported complaint.Additional product testing has been performed since initial confirmation to expand the scope of lot numbers impacted as well as determine impact on performance.Complaint history review: initial complaint history review identified two tickets documented for ic failure with ec 4457 and are related to the reported issue; initial investigation for these two tickets identified no product deficiency.Subsequent complaint searches performed as part of the risk assessment process and (b)(4) capa have identified 35 tickets that are related to this issue.Based on the results of the investigation elements a product deficiency for abbott realtime hiv-1 amp rgt kit-row (list 02g31-010) lot 476139 was identified.Specifically, based on in-house product evaluation file sample testing that verified the issue reported by the customer.(b)(4) capa and (b)(4) identified a second lot of amplification reagents exhibiting similar performance, and, as a result, u.S.Abbott realtime hiv-1 amplification reagent kit, list 06l18-90, lot 475933 was also included in the scope of the investigation.As the root cause investigation is currently in progress, an interim action was initiated to evaluate whether additional lots were demonstrating the same performance issues as lots 476139 and 475933.Based on the evaluation criteria of the interim action, a statistical process control (spc) analysis and timeframe of lots that are included in this risk and shared components a total of 14 lots have been included.Abbott molecular determined that a field corrective action (fa-am-aug2017-224) should be taken and reported this issue per 21 cfr 806 to the fda (b)(6) district recall coordinator on august 1, 2017 (report number 3005248192-08/01/2017-001-c).The field corrective action was in the form of a field correction recall letter / urgent field safety notice, dated august 8, 2017 instructing customers to discontinue use of and discard all remaining inventory of all realtime hiv-1 quantitative lots listed in the letter.Also included in the letter were considerations for healthcare providers as guidance for retesting hiv-1 viral loads of potentially impacted hiv patients.Date of this event is (b)(6) 2017, which is the date the ticket was received.
 
Event or Problem Description
The abbott realtime hiv-1 assay is an in vitro reverse transcription polymerase chain reaction (rt-pcr) assay for the quantitation of (b)(6) on the automated m2000 system in human plasma from (b)(6) infected individuals over the range of 40 to 10,000,000 copies/ml.The abbott realtime hiv-1 assay is intended for use in conjunction with clinical presentation and other laboratory markers for disease prognosis and for use as an aid in assessing viral response to antiretroviral treatment as measured by changes in plasma (b)(6) levels.This assay is not intended to be used as a donor screening test for (b)(6) or as a diagnostic test to confirm the presence of (b)(6) infection.A customer complaint from (b)(6) was received indicating receipt of error codes related to non-us list 02g31-010 lot 476139.Based on review of customer run data and internal testing, the complaint investigation was confirmed for reduced pcr amplification efficiency, which presents as delayed threshold cycle (ct) values and reduced max ratio (mr) values.On june 22, 2017, this customer complaint from the united states was received.Customer indicated error code generation and non-abbott controls coming up not detected (expected result = 200 copies / ml).Based on in-process cause investigation and this complaint, a second lot of amplification reagents was determined to be exhibiting similar performance.As a result, u.S.Abbott realtime hiv-1 amplification reagent kit, list 06l18-90, lot 475933 was included in the scope.The medical evaluation concluded that the hazards and harms identified in the malfunction risk evaluation (360-risk) would not be expected to cause or contribute to death or serious injury.This complaint is reported in accordance with 21 cfr 803 on the basis that a reportable (21 cfr 806) field action, fa-am-aug2017-224, was taken to address the issue.
 
Additional Manufacturer Narrative
Labeled for single use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABBOTT REALTIME HIV-1 AMP REAGENT KIT
Common Device Name
TEST, HIV DETECTION
Manufacturer (Section D)
ABBOTT MOLECULAR, INC.
1300 e. touhy ave.
des plaines IL 60018 3315
MDR Report Key6779184
Report Number3005248192-2017-00003
Device Sequence Number10797424
Product Code MZF
UDI-Device Identifier00884999023949
UDI-Public(01)00884999023949(10)475933(17)180403
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
BP060002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date04/03/2018
Device Model Number6L18-90
Device Catalogue Number06L18-90
Device Lot Number475933
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 07/27/2017
Supplement Date Received by Manufacturer08/11/2017
Initial Report FDA Received Date08/09/2017
Supplement Report FDA Received Date08/24/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-