• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC; ANALYZER, PACEMAKER GENERATOR FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC; ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Medical Device Problem Codes Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported an error message appeared during implant indicating the programmed value or programmed mode could not be accepted and to exit the pop-up box and select a new parameter.This occurred after the user tried to assess lv (left ventricular) lead placement, while testing lv tip to ring configuration.As a result, a pacing mode or rate to analyze the lead could not be set.The user then exited the pop-up and tried to re-select, but the system locked, would not allow them to advance, and wouldn't launch after.The application was forced closed, and the base station power was disconnected and reconnected, which did not work.Finally, a soft reset of the tablet worked.A different programmer/analyzer model was used to finish the case.The mobile programmer application remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Analysis summary: the reported out of specification issues with the mobile programmer application were confirmed.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported an error message appeared during implant indicating the programmed value or programmed mode could not be accepted and to exit the pop-up box and select a new parameter.This occurred after the user tried to assess lv (left ventricular) lead placement, while testing lv tip to ring configuration.As a result, a pacing mode or rate to analyze the lead could not be set.The user then exited the pop-up and tried to re-select, but the system locked, would not allow them to advance, and wouldn't launch after.The application was forced closed, and the base station power was disconnected and reconnected, which did not work.Finally, a soft reset of the tablet worked.A different programmer/analyzer model was used to finish the case.The mobile programmer application remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK SMARTSYNC
Common Device Name
ANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key6780194
Report Number3004593495-2017-00397
Device Sequence Number999711
Product Code DTC
UDI-Device Identifier00643169868540
UDI-Public00643169868540
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K163008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24970A
Device Catalogue Number24970A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/07/2017
Supplement Date Received by Manufacturer10/11/2017
Initial Report FDA Received Date08/09/2017
Supplement Report FDA Received Date12/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-