| Model Number |
D334TRM |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/03/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical device: 4396-88 lead, implanted: (b)(6) 2013.Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.
|
| |
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited premature battery depletion.The device was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|