Model Number 1506-8600-000
Device Problem Defective Alarm (1014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/12/2017
Event Type  malfunction  
Manufacturer Narrative
Replaced power management board to fix the reported issue.Patient information could not be obtained after multiple attempts.Attempts were made as follows: 8/7/2017 phone call, 8/8/2017 phone call, 8/9/2017 phone call.Unique device identifier: (b)(4).
 
Event Description
The hospital reported that, during patient use, the vent stopped and then restarted without alarming.There was no reported patient injury.
 
Manufacturer Narrative
According to gehc engineering review, the power management board failure impacted display only.Review of the logs revealed there was no loss of ventilation.Therefore, the initial mdr was not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESCAPE R860
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
540 w. northwest highway
barrington, IL 
MDR Report Key6784055
MDR Text Key82601386
Report Number2112667-2017-01507
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1506-8600-000
Other Device ID NumberSEE BLOCK H10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N