Model Number 1506-8600-000 |
Device Problem
Defective Alarm (1014)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/12/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Replaced power management board to fix the reported issue.Patient information could not be obtained after multiple attempts.Attempts were made as follows: 8/7/2017 phone call, 8/8/2017 phone call, 8/9/2017 phone call.Unique device identifier: (b)(4).
|
|
Event Description
|
The hospital reported that, during patient use, the vent stopped and then restarted without alarming.There was no reported patient injury.
|
|
Manufacturer Narrative
|
According to gehc engineering review, the power management board failure impacted display only.Review of the logs revealed there was no loss of ventilation.Therefore, the initial mdr was not reportable.
|
|