| Model Number |
019-540 |
| Medical Device Problem Codes |
Device Issue (2379); Connection Problem (2900)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
07/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was discarded at hospital and is not returning to a and e medical for evaluation.A and e medical quality system documentation (complaints, nonconformces, and device history records have been reviewed).These records do not indicate any trends pertaining to this device and the alleged complaint.There are no other complaints within the a and e medical system.This complaint would appear to be an isolated event.The patient was noted as being paced with this device prior to movement of the patient.It is possible the connector could have been damaged while transporting patient from intensive care unit to operating room.A and e medical will continue to monitor its quality system.
|
| |
|
Event or Problem Description
|
|
Disposable myo lead cable came unplugged from the pacing box.The patient coded in the cvicu and in the cvor.The connector circumference of the plug was observed having a fracture.The fracture created a "flimsy" interface with the pacing box.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was discarded at hospital and is not returning to a&e medical for evaluation.A&e medical quality system documentation (complaints, nonconformces, and device history records have been reviewed).These records do not indicate any trends pertaining to this device and the alleged complaint.There are no other complaints within the a&e medical system.This complaint would appear to be an isolated event.The patient was noted as being paced with this device prior to movement of the patient.It is possible the connector could have been damaged while transporting patient from intensive care unit to operating room.A&e medical will continue to monitor its quality system.(b)(4).
|
| |
|
Event or Problem Description
|
|
Disposable myo lead cable came unplugged from the pacing box.The patient coded in the cvicu and in the cvor.The connector circumference of the plug was observed having a fracture.The fracture created a "flimsy" interface with the pacing box.
|
| |
|