• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6)2017, the reporter contacted animas, alleging a button/keypad (tactile changes/unresponsive) issue.It was reported that the up arrow, down arrow, and ok/enter buttons were under-responsive.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue has the ability to result in inadvertent or incorrect insulin delivery or the inability to use the pump.
 
Additional Manufacturer Narrative
Follow-up #1: date of submission 10/10/2017 device evaluation: the device has been returned and evaluated by product analysis on 09/25/2017 with the following findings: during investigation, the up arrow, down arrow, and ok keypad buttons were unresponsive.The keypad was removed to find no damage to the button contacts or keypad flex cable.The pump was opened to find moisture corrosion on the keypad connector.No moisture was found in the battery compartment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key6784945
Report Number2531779-2017-18007
Device Sequence Number139216
Product Code OYC
UDI-Device Identifier10840406100020
UDI-Public0110840406100020
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 07/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2017
Was the Report Sent to FDA? Yes
Device Age15 MO
Initial Date Received by Manufacturer 07/30/2017
Supplement Date Received by Manufacturer07/30/2017
Initial Report FDA Received Date08/10/2017
Supplement Report FDA Received Date10/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age18 YR
-
-