| Model Number |
HT70 |
| Medical Device Problem Code |
Failure to Calibrate (2440)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the ht70 ventilator failed oxygen (o2) sensor calibration.There was no patient involvement at the time of the event.Medtronic was not authorized to evaluate/repair the device as the product is not in medtronic control.The repair will be completed at a non-medtronic location.The reported complaint could not be confirmed without the product return.Should new information become available or if the product is returned to the manufacturer for investigation the complaint will be re-evaluated.
|
| |
|
Additional Manufacturer Narrative
|
|
The service provider checked the ventilator and found the single board computer, controller printed circuit board, handle broken and will need to be replaced.If information is provided in the future, a supplemental report will be issued.
|
| |
|