Model Number MMT-751NAS
Device Problems Moisture Damage (1405); Improper or Incorrect Procedure or Method (2017); Moisture or Humidity Problem (2986)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/15/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call moisture damage and problems with the insulin pump.Customer's blood glucose level was 200 mg/dl.Troubleshooting for moisture damage was performed.Customer advised they fell into the swimming pool while wearing the insulin pump.Customer advised the insulin pump would short out because of the fluid inside of the insulin pump.Customer was advised to discontinue use of the device and revert to a backup plan.Customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
Findings: compromised force sensor system alarm during the prime test at 4.0 pounds due to faulty sensor resistor.Pump passed the operating current measurement, self test, unexpected alarm error test, rewind, basic occlusion test and displacement test.Unable to perform the occlusion test and no delivery test due to compromised force sensor system alarm.Moisture damage found on the lcd board.Pump had cracked reservoir tube lip and minor scratched display window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6786702
MDR Text Key82719006
Report Number2032227-2017-35976
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503717
UDI-Public(01)00643169503717
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 10/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAS
Device Catalogue NumberMMT-751NAS
Device Lot NumberA4751NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/15/2017
Initial Date FDA Received08/10/2017
Supplement Dates Manufacturer Received10/02/2017
Supplement Dates FDA Received10/06/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age26 YR
Patient Weight180
-
-