Findings: unit power up properly after battery installation.No blank display noted.Unit was received with normal operating currents and no unexpected off no power, weak battery, bad battery, low battery, battery out limit and failed battery test alarms during testing.Unit passed self-test, unexpected alarm error test, rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.Unit had minor scratched lcd window, cracked battery tube threads and cracked reservoir tube lip.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
|