COOK IRELAND LTD ECHOTIP ULTRA ENDOBRONCHIAL HD ULTRASOUND NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G53408
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) # k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.1 x echo-hd-22-ebus-p was returned to cirl for evaluation.Upon evaluation of the returned device the following was noted: the syringe was attached to the handle on return.The stylet returned with the device.Part of the needle was returned in a container as per description of events the needle was cut the sheath extender was fine.The handle could not be moved forward.The broken part of the needle is kinked distally.The needle was removed from the device.The difficulty with the handle was due to the needle being cut by the user.The customer complaint is considered to be confirmed as needle was kinked distally.A possible root cause could be due to a tortuous anatomy and excessive force being applied to the entry point cause the needle to bend.Prior to distribution, all echo-hd-22-ebus-p devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.The notes section of the instructions for use which accompanies this device instructs the user to ¿visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use.¿ from the information provided, no adverse effects to the patient have been reported as a result of this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
Distal end of needle kinked and was clipped off with a forceps."as per complaint form": distal end of needle kinked after punction and was clipped off with a forceps in order to fit needle back in sheath.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP ULTRA ENDOBRONCHIAL HD ULTRASOUND NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key6787654
Report Number3001845648-2017-00323
Device Sequence Number121127
Product Code FCG
UDI-Device Identifier00827002534087
UDI-Public(01)00827002534087(17)200502(10)C1352000
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberG53408
Device Catalogue NumberECHO-HD-22-EBUS-P
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/11/2017
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2017
Initial Report FDA Received Date08/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-