MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) PERCUTANEOUS REFERENCE CROSS PIN FRAME; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) PERCUTANEOUS REFERENCE CROSS PIN FRAME; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9732353
Medical Device Problem Codes Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device lot number, or serial number, unavailable.Device manufacturing date is dependent on lot number/serial number, therefore, unavailable.No parts have been received by the manufacturer for evaluation.
 
Event or Problem Description
A medtronic representative reported that, while in a spinal fusion, the perc pin was loose resulting in slight movement of the reference frame.The site elected to continue with the frame and ensuring that the frame would not move.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no impact on patient outcome.No additional information was provided.
 
Additional Manufacturer Narrative
Additional information: device manufacture date and lot number provided.The suspect frame was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUTANEOUS REFERENCE CROSS PIN FRAME
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key6788797
Report Number1723170-2017-03288
Device Sequence Number121710
Product Code OLO
UDI-Device Identifier00613994151179
UDI-Public00613994151179
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K131425
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date (Section B) 09/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number9732353
Device Lot Number150708
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2017
Initial Date Received by Manufacturer 07/14/2017
Supplement Date Received by Manufacturer08/24/2017
Initial Report FDA Received Date08/11/2017
Supplement Report FDA Received Date09/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight110
-
-