DATASCOPE MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0998-00-3023-53
Device Problems Loss of Power (1475); Low Battery (2584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/18/2017
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date.The field service engineer (fse) evaluated the intra-aortic balloon pump (iabp) and was able to duplicate the reported issue.The fse noted that the batteries were last replaced (b)(6) 2015.He performed a battery run-time test after the customer plugged in the iabp overnight.The batteries only ran for 30 minutes with the patient simulator and 40 cc balloon attached.The fse replaced the expired batteries and performed all functional and electrical safety tests.The iabp successfully passed all tests.The fse informed the customer to fully charge batteries overnight before transporting patient on the iabp.The iabp was returned to the customer and cleared for clinical use.
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) failed during patient transport and it shutdown inexplicably on its battery.The customer stated that the iabp stopped running on back up power during patient transport to another hospital and the low battery alarm generated.There was no patient injury or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key6788905
MDR Text Key82770035
Report Number2249723-2017-00268
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Device AgeYR
Date Manufacturer Received07/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-