Model Number MMT-551NAS
Device Problems Component Falling (1105); Improper or Incorrect Procedure or Method (2017); Material Protrusion/Extrusion (2979)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/17/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call that the device was dropped and drive support cap was sticking out.Customer's blood glucose was unknown.Customer was advised to discontinue use of the device and revert to a back-up plan per health care professionals¿ instructions.Customer was advised that the insulin pump will be replaced.Customer agreed to return the device for analysis.
 
Manufacturer Narrative
Findings: unit had minor scratched lcd window, cracked battery tube threads, cracked reservoir tube lip, reservoir tube cracked and stripped battery cap coin slot (p).Drive support disk was inspected and no anomaly was noted.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551NAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6789330
MDR Text Key82700212
Report Number2032227-2017-36235
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503663
UDI-Public(01)00643169503663
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 10/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-551NAS
Device Catalogue NumberMMT-551NAS
Device Lot NumberA4551NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received10/06/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age58 YR
Patient Weight195
-
-