Catalog Number 190766
Device Problem Fire (1245)
Patient Problem No Patient Involvement (2645)
Event Date 07/18/2017
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A biomedical technician at a user facility reported to technical support that a 2008t hemodialysis (hd) machine had valve 103 caught fire during bleach disinfect.No patient was connected to the machine at the time of the incident.Therefore, no patient was involved.Follow-up confirmed the use of hospital grade ground fault circuit interrupter (gfci) outlets at the user facility.The biomedical technician replaced the bibag hydraulics assembly to resolve the issue.Following the parts replacement, the system was restored to full functionality.Functional testing performed by the biomedical technician confirmed the system was operating properly.The unit has been returned to service at the user facility without a recurrence of the event as reported.No parts were available to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
The device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed by a fresenius regional equipment specialist (res) and no parts were returned for failure analysis.The biomedical technician indicated that the bibag hydraulics assembly was replaced which resolved the issue.The unit has been returned to service at the user facility without a recurrence of the event as reported.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process which could be associated with the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM W/BIBAG
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6791046
MDR Text Key82564219
Report Number2937457-2017-00715
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100910
UDI-Public00840861100910
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Biomedical Engineer
Device Catalogue Number190766
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Initial Date Manufacturer Received 07/18/2017
Initial Date FDA Received08/14/2017
Supplement Dates Manufacturer Received09/02/2017
Supplement Dates FDA Received09/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-