Model Number 4415 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/22/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The fsr determined that a defective cpg temperature sensor was the cause of the inaccurate readings.The sensor was replaced.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.This complaint is related to (b)(4)/ medwatch #1828100-2017-00365.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the heater cooler was displaying an inaccurate cardioplegia (cpg) water temperature, at 50.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) measured the temperature sensor and found it to be electrically open.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|