The investigation determined that lower than expected vitros amon results were obtained when testing non-vitros quality controls on a vitros 5600 integrated system.The assignable cause of this event was instrument related.The results of a diagnostic within run vitros amon precision test was not acceptable indicating an instrument related issue.After service actions were completed to clean the evaporation caps and adjusting the slide align guides, the precision was acceptable indicating service actions had returned the analyzer to acceptable performance.The customer continued to obtain unacceptable amon results and an issue was identified involving the slide supply humidity affecting quality control results for multiple assays.After the ortho fe replaced the humidity plate, post-service quality control testing was acceptable indicating the actions have resolved the issue.Since unacceptable performance was observed for multiple reagents including vitros amon and acceptable performance was observed after the service actions using the current reagent lot, vitros amon lot 1017-0245-0578 can be ruled out as a contributing factor.
|