| Catalog Number |
309.490 |
| Medical Device Problem Codes |
Break (1069); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device used for treatment, not diagnosis.Patient information not available for reporting.Implant and explant dates: device is an instrument and is not implanted/explanted.Device is expected to be returned to synthes manufacturer for evaluation /investigation, but has yet to be received.Initial reporters phone number: (b)(6).Device history records review was conducted.The report indicates that the: part #09.490/ lot #9788362.Manufacturing location: (b)(4).Manufacturing date: 28.Dec.2015.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes on an event in (b)(6) as follows: it was reported that during an intervention on (b)(6) 2017 two extraction bolts damaged.One extraction bolt broke.The other bolt had a screw stuck inside and the technician did not manage do get the screw out.There was no impact on the patient and the patient is fine.The surgery was not prolonged.Only one extraction bolt was affected, the one that is broken.Additionally, a hollow reamer had a screw stuck inside and the technician was not able to get the screw out.This complaint involves 1 part.This report is 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been received and the product evaluation is in progress.No conclusion can be drawn.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Further it was clarified that only one extraction bolt is affected.Extraction bolt has fragment left inside it and the hollow reamer has broken thread.This is report 1 of 2 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Customer quality conducted an investigation of the returned devices.Received parts: 1 x 309.490 / extract-bolt f/scr ø4.5+5 / lot no 9788362.1 x 309.035 / hollow-reamer-compl anticlockwise cutting / lot no 2029755.As received condition: 309.490: the part shows slightly scratches on surface.Furthermore we found a broken screwhead, stucked inside the inner-threaded hole of device.The broken part is missing.309.035: the surface shows slightly mechanical damages and the outside-thread is broken.The broken part is missing.Conclusion for 309.490: our investigation has shown that there are slightly scratches on surface.Furthermore we found a broken screwhead, stucked inside the inner-threaded hole of device.The broken part is missing.The manufacturing review for does show that the production procedure was according to the specifications and there were no issues that would contribute to this complaint condition.Unfortunately we only have limited information in the complaint description and cannot confirm how this happened.Because of the damage and without broken part the complaint relevant dimensions cannot be checked for dimensional accuracy.We do suppose that the cause of the breakage is the result of a mechanical overload situation during use.Overall conclusion: based on the investigation results we conclude that the cause of failure is not due to any manufacturing non-conformances.Corrected data: description clarification, concomitant devices.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clarification of event: it is the extraction bolt that has a broken screw head stuck inside of it.The hollow reamer is broken.Concomitant devices reported: screw (part and lot unknown, quantity 1).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: device available for evaluation: returned to manufacturer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes europe reports an event in (b)(6) as follows: it was reported during an unknown procedure on (b)(6) 2017, and extraction bolt for 4.5mm screws broke and the thread of a hollow reamer complete for 3.5mm & 4.0mm screws is broken and has a screw stuck inside of it.Surgery was completed successfully with no delay.Concomitant devices reported: screw (part number unknown, lot number unknown, quantity 1).This report is for one (1) extraction bolt for 4.5mm screws.This is report 1 of 2 for (b)(4).
|
| |
|
Search Alerts/Recalls
|