Model Number 33600030
Device Problem Device Inoperable (1663)
Patient Problem Visual Disturbances (2140)
Event Date 07/18/2017
Event Type  Injury  
Manufacturer Narrative
Neither the device nor applicable imaging films were returned to the manufacturer for evaluation, therefore, the cause of the event cannot be determined.
 
Event Description
It was reported that the patient underwent a total ankle replacement surgery.Allegedly, the adjustment block screw was lodged in adjustment block and could not be backed out to lock into the coronal guide.The surgeon postponed the case after the patient was already put under anesthesia.No additional patient complications were reported.
 
Manufacturer Narrative
Visual examination of the returned part does not show any overall gross deformation.Upon an initial functional review of the devices, some of the lock outs seemed jammed but were able to be released without any additional aid of lubrication and/or power, and then the adjustment block knobs functioned as intended.Also, the functional evaluation included a review of the coronal guide lockout, in which a coronal guide was able to be locked into place.Therefore, no failure was detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY®
Type of Device
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
38002
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
38002
Manufacturer Contact
matthew parrish
1023 cherry road
901451-631
MDR Report Key6793026
MDR Text Key82653126
Report Number1043534-2017-00084
Device Sequence Number1
Product Code HSN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123954
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number33600030
Device Lot Number1582249
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2017
Is the Reporter a Health Professional? No
Distributor Facility Aware Date07/18/2017
Event Location Hospital
Date Manufacturer Received09/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-