| Model Number |
1 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/11/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: unknown.Incident description: (b)(4) incident reported by email by clinical procurement.Details of the complaint listed as "the blue pad from the tip of the grasper came off".Patient status: unknown.Type of intervention: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the event unit was returned for evaluation.Visual inspection confirmed the complainant's experience of pad detachment.Based on the condition of the returned unit, it is likely that the reported event was caused during manufacturing, either from a lack of adhesive or from repositioning or readjusting the pad during curing.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.
|
| |
|