• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SILICONE CHANNEL DRAIN WITHOUT TROCAR, 19 FR., ROUND HUBLESS, FULL FLUTED; DRAIN CHANNEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SILICONE CHANNEL DRAIN WITHOUT TROCAR, 19 FR., ROUND HUBLESS, FULL FLUTED; DRAIN CHANNEL Back to Search Results
Catalog Number 072190
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the drain appeared to be the incorrect size.There was no patient involvement.
 
Manufacturer Narrative
Received 19 unopened channel drains within 2 case boxes.The reported event was confirmed as manufacturing related.Per the visual inspection, there were no obvious defects on all the returned 19 channel drains found.However, per dimensional evaluation, the 19 samples were measured with a french scale and with the electronic measurement vision system and the all samples were confirmed to correspond to one piece channel drain 24fr instead of 19 fr.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "reorder # v072230, drain size:19 fr., drain shape: fluting:round hubless, trocar:full depth: ¿, indicator: dot.V072234, 24 fr., round hubless, full, ¿, dot.For use with 100cc, 200cc, 400cc, and 3-spring evacuators.Indications: round silicone channel drains are indicated for use with selected evacuators for closed wound drainage following head and neck, orthopedic, abdominal, ent, ob/gyn, plastic, neurosurgery, thoracic and cardiovascular (channel drains only) procedures.Complications: complications which may result from the use of this suction drainage system include the risks associated with methods utilized in the surgical procedure, as well as the patients degree of intolerance to any foreign object in the body.The advantages of wound drainage, particularly closed system drainage, are lost if an air-tight seal between the drain and the skin where the drain emerges is not achieved, or if the drain is allowed to become occluded or if the reservoir is not activated properly, doesn¿t function properly or is not monitored.Evacuators should be emptied and re-activated when required per hospital protocol.In the event an air-tight seal is not achieved, the reservoir will rapidly fill with air from the leak; subsequent drainage to the reservoir will occur only if allowed by gravity and wound exudate forcing the flow.Entry into the reservoir is allowed only by displacement of air in the reservoir by wound exudate flow.In this displacement process, air reflux from the reservoir to the wound can occur and increase the likelihood of back-contamination across the anti-reflux valve.In the event of drain occlusion by fibrin, clots, or other particulate matter, all wound drainage via the drain ceases.If the reservoir is not emptied when it is full, equilibrium between the drain and reservoir at wound pressure will ultimately occur and drainage from the wound site will cease.When the reservoir and drain are at the same pressure and the reservoir is full of fluid, the likelihood of back-contamination across the anti-reflux valve is increased.When used to drain the pleural cavity in the presence of an air leak, drains must be attached to an appropriate pleural cavity drainage system to prevent tension pneumothorax.Drain placement: the surgeon should irrigate the wound with sterile fluid, then suction the irrigating fluid and gross debris from the operative site.Tubes should lie flat and in line with the anticipated skin exit.To facilitate later removal by manual traction, the tubing should not be curled, pinched, or sutured internally.Positioning of the drain in the body cavity, as well as the number of drains indicated, should be determined by the operating surgeon.Drain tubing should be placed within the wound by approximating the areas of critical fluid collection.Care must be taken to ensure that all drain perforations or channels lie completely within the wound or cavity to be drained.Taping or a triple loop suture (around and not through the tubing) will aid in preventing accidental drain displacement.Deep drainage is best accomplished by using one or more drains for each level of tissue.Each level should be evacuated by a separate vacuum source.Care must be exercised to avoid damage to the drain (see warnings).The tubing should be repeatedly checked during closure for free motion to avoid breakage and/or fragment retention within the wound." (b)(4).
 
Event Description
It was reported that the drain appeared to be the incorrect size.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILICONE CHANNEL DRAIN WITHOUT TROCAR, 19 FR., ROUND HUBLESS, FULL FLUTED
Type of Device
DRAIN CHANNEL
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6793512
MDR Text Key82887385
Report Number1018233-2017-04225
Device Sequence Number1
Product Code GBX
UDI-Device Identifier00801741051616
UDI-Public(01)00801741051616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Catalogue Number072190
Device Lot NumberNGAZ3014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/24/2017
Initial Date FDA Received08/14/2017
Supplement Dates Manufacturer Received10/10/2017
Supplement Dates FDA Received10/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-