| Model Number |
MT20649 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2017 that on (b)(6) 2017 the receiver had intermittent audio output.No additional event or patient information is available.The receiver has been received for evaluation. a follow-up report will be submitted once the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Receiver was charged and booted up.Functional testing was performed and the reported fault could not be reproduced and there was no failure detected.Global communication tool software test was performed and passed.A review of the downloaded data log did not find any errors related to the customer complaint.A manual test was performed and speaker sounded.A manual drop test for intermittency was performed and the test passed.The receiver case was opened for further evaluation.A visual interior inspection was performed and the inspection passed.A speaker resistance test was performed and passed.The device was determined to be operating within the required specifications without malfunction.The reported event of an intermittent audio output was not confirmed.A root cause could not be determined.
|
| |
|