• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN NORTH HAVEN - MFG VERISET; AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN NORTH HAVEN - MFG VERISET; AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED Back to Search Results
Model Number HP0510E
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/18/2017
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a laparoscopic cholecystectomy procedure, the device mesh was damaged, it was flaky and crumbling on touch.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of the device.The visual inspection of the sample noted that one of the foils pouches was received from the account opened without the tyvek pouch.The patch was broken and degraded.The five sealed samples were opened and inspected; no abnormalities were noted.The foil pouches were inspected with light assisted microscopy with no abnormalities noted.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures and a review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERISET
Common Device Name
AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED
Manufacturer (Section D)
COVIDIEN NORTH HAVEN - MFG
195 mcdermott road
north haven CT 06473
MDR Report Key6794525
Report Number1219930-2017-06419
Device Sequence Number11196360
Product Code LMG
UDI-Device Identifier10884521082557
UDI-Public10884521082557
Combination Product (Y/N)N
Initial Reporter CountryEI
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/11/2018
Device Model NumberHP0510E
Device Catalogue NumberHP0510E
Device Lot NumberN7D0325UX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2017
Initial Date Received by Manufacturer 07/18/2017
Supplement Date Received by Manufacturer10/24/2017
Initial Report FDA Received Date08/15/2017
Supplement Report FDA Received Date11/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age27 YR
-
-