| Catalog Number |
C-TQTS-1200 |
| Health Effect - Clinical Code |
Nervous System Injury (2689)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being filed after the associated complaint was reviewed under remediation protocol_cap008, compliant/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
|
| |
|
Event or Problem Description
|
|
The markings on the chest tube does not account for the 3 millimeter part of the catheter where the holes are.The user states this creates the potential for the catheter being placed too far.The user facility noted incidents where the drains have been placed too far.In one case it was placed to far in a preterm infant where diaphragmatic paralysis due to phrenic nerve palsy (captured on report number 1820334-2016-00437).
|
| |
|