• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC THAL-QUICK CHEST TUBE SET; GBX CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC THAL-QUICK CHEST TUBE SET; GBX CATHETER, IRRIGATION Back to Search Results
Catalog Number C-TQTS-1200
Health Effect - Clinical Code Nervous System Injury (2689)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol_cap008, compliant/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
The markings on the chest tube does not account for the 3 millimeter part of the catheter where the holes are.The user states this creates the potential for the catheter being placed too far.The user facility noted incidents where the drains have been placed too far.In one case it was placed to far in a preterm infant where diaphragmatic paralysis due to phrenic nerve palsy (captured on report number 1820334-2016-00437).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THAL-QUICK CHEST TUBE SET
Common Device Name
GBX CATHETER, IRRIGATION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6795516
Report Number1820334-2017-02421
Device Sequence Number771930
Product Code GBX
UDI-Device Identifier00827002051614
UDI-Public(01)00827002051614(17)181104(10)6353821
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue NumberC-TQTS-1200
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2017
Initial Report FDA Received Date08/15/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-