| Model Number |
E SERIES |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional procode = dro.Zoll medical corporation evaluated the device and the reported malfunction was observed but not duplicated.The device was put through extensive testing without duplicating the malfunction.The device's system board frame, connector interlock, and a system interconnection flex cable were replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed "defib fault 72" and "pacer fault 116" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|