| Model Number |
E SERIES |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to the lithium battery on the system board.The device operator's guide instructs users to replace the lithium battery on the device every 5 years.This device is approximately 9 years old.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|