• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HAEMONETICS CORP CELL SAVER® ELITE® AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS CORP CELL SAVER® ELITE® AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE Back to Search Results
Model Number CSE-P-225
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 07/18/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based upon the clinical information available it is not possible to determine whether the patient's critical condition was due in part or in entirety as a result of the leaking disposable bowls.The bowls were returned for evaluation, a visual inspection confirmed the presence of cracks along the base of the bowl.The base diameter of the bowl was measured on each unit, all four units were found to meet dimensional specifications.The bowls were installed into a centrifuge chuck without problem.Submersion leak testing was performed, and a leak was confirmed in three of the four returned bowls.A root cause of the bowl failure has not been established at this time, it is not possible to determine at which point during the product lifecycle the cracks developed along the bowl weld, it is possible that the bowls were damaged at some point after being released from manufacturing prior to use, either during product handling while shipping or during product handling during installation and procedure setup.Haemonetics attempted to reproduce the failure mode for cracks forming along the base weld of the bowl as well as for a leak at the base weld, it was not possible to reproduce the damages observed in the returned samples through the manufacturing process.The manufacturing process employs a 100% dynamic leak test for these products, the gross leaks found on the returned products would have been detected during the dynamic leak test.The damages which resulted with leaking products were not caused by any observable error or deviation in the manufacturing process.
 
Event or Problem Description
The event occurred (b)(6) 2017 during a surgical procedure to repair an abdominal aortic aneurysm.The patient required multiple units of rbcs during the surgery and post op due to leakage of disposable bowls into the well of the cell saver machine which prompted a shutdown of the machine.This occurred with three separate machines.The disposables appeared to have a leak at the bottom of the bowl.As each machine became disabled, another system was set up with a new disposable.The problem kept repeating until all machines were not functional and 5 liters of blood collected in a reservoir that was unable to be processed.A belmont rapid infuser was used with bank blood to replace the blood loss.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER® ELITE® AUTOTRANSFUSION SYSTEM
Common Device Name
CELL SAVER ELITE
Manufacturer (Section D)
HAEMONETICS CORP
400 wood rd
braintree MA 02184
MDR Report Key6797647
Report Number1219343-2017-00016
Device Sequence Number590479
Product Code CAC
UDI-Device Identifier10812747016544
UDI-Public(01)10812747016544(17)200411(10)0417067
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K101907
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date04/11/2020
Device Model NumberCSE-P-225
Device Catalogue NumberCSE-P-225
Device Lot Number0417067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2017
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/19/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2017
Initial Report FDA Received Date08/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient Weight88
-
-