| Model Number |
286735350 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Review of the device history record identified no issues during the manufacturing or release of the product that could be attributed to the problem reported by the customer.The product was released accomplishing all quality requirements.A follow up report will be filed upon completion of the investigation.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Orthokit specialist checked devices and found next damages: code 286720000 - the thread damage code 286735350 - had the working part of the screwdriver code 286735400 - the failure of the slot code 286745550 - the working part was broken.The date of inspection is (b)(6) 2017 and we accepted this date as date of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The x25 approximator was returned for evaluation.Visual inspection revealed that the distal tip of the driver was missing.Optical imaging found evidence of a torsional shear overloading.Review of the device history record identified no issues during the manufacturing or release of the product that could have contributed to the problem reported by the customer.The product was released accomplishing all quality requirements.A review of the complaint trend analysis was performed and no systemic trend was identified as a result of the analysis.The root cause cannot be positively determined however the failure is mostly due to unanticipated torsional loads.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends have been observed, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|