MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD AUTO SUTURE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD AUTO SUTURE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 24055-
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/18/2017
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, both devices broke on the patient.No patient injury has been noted.Medical intervention was required.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of the device.The visual inspection of the returned product noted; the spiked trocar was received.The circular and envelope seals were intact.The flanges of the anchors were skewed and broken.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the reported condition may be due to insertion or removal of the port without the anchor tabs being fully closed.Under these circumstances, the tabs may be exposed to excessive force and therefore break.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTO SUTURE
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key6798538
Report Number9612501-2017-05970
Device Sequence Number16403899
Product Code GCJ
UDI-Device Identifier10884521080829
UDI-Public10884521080829
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Model Number24055-
Device Catalogue Number24055-
Device Lot NumberJ5C0070X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2017
Initial Date Received by Manufacturer 07/19/2017
Supplement Date Received by Manufacturer10/10/2017
Initial Report FDA Received Date08/16/2017
Supplement Report FDA Received Date10/31/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-