| Model Number |
E2781R-28ASP |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/09/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One used e2781r-28asp spatula was received, and an evaluation of the sample confirmed an issue with the insulation.Visual inspection found the device was melted around the aspiration holes.The issue caused the device to fail hipot testing.The melting occurred as a result of arcing, which can be due to simultaneously activating the suction function and electrical current.It can also occur if the electrode is activated while retracted in the sheath.The investigation identified the root cause of the reported event to be user error.The instructions for use included with this device cautions users not to simultaneously activate the suction/irrigation and electrosurgical current.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a procedure, energy emitted from the tip of the shaft instead of the spatula when activated near the urinary tract.A new device was used to continue the procedure.No patient injury occurred.During medtronic's initial evaluation of the received device, it was found that part of the insulation around the aspiration holes was melted.The device also failed hipot testing around the melted insulation.
|
| |
|