Model Number ENBF3216C166E
Medical Device Problem Codes Leak/Splash (1354); Migration or Expulsion of Device (1395)
Health Effect - Clinical Codes Pain (1994); Rupture (2208)
Date of Event 07/20/2017
Type of Reportable Event Serious Injury
Event or Problem Description
An endurant ii stent graft system was inserted in a in a patient for the endovascular treatment of a 58.65mm diameter abdominal aortic aneurysm.It was reported that the patient presented emergently in the emergency room on with abdominal/back pain.A ruptured infra-renal aneurysm was discovered via ct scan.The patient was transferred directly into surgery where the physician explanted the endurant stent graft.Per the physician the cause of the event was due to persistent type ii endoleak resulting in proximal neck dilatation leading to proximal graft migration of 5-10mm and an eventual type 1a endoleak which was noted at the time of explant surgery.No additional clinical sequelae were reported and the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6799815
Report Number2953200-2017-01305
Device Sequence Number16529075
Product Code MIH
UDI-Device Identifier00613994762429
UDI-Public00613994762429
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/06/2013
Device Model NumberENBF3216C166E
Device Catalogue NumberENBF3216C166E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2017
Initial Report FDA Received Date08/17/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
-
-