| Model Number |
H74939171120610 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the iliac vein.A 12.0 x 60, 135cm mustang balloon catheter was advanced for dilatation.However, the balloon ruptured during procedure.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device lot number corrected from 0020322602 to 20131245.Device expiration date corrected from 02/22/2020 to 01/06/2020.Device manufactured date corrected from 02/24/2017 to 01/10/2017.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was further reported that the balloon was inflated twice.The first inflation was done at 10 atmospheres; however, upon the second inflation, the balloon ruptured.The device was completely removed all together with the guide catheter and the procedure was completed with another device.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: a mustang balloon device was received for analysis.A visual examination confirmed that the balloon had been subjected to positive pressure.It was observed that the balloon was separated into two separate pieces; both pieces were fully attached to the balloon shaft.A visual examination identified a complete circumferential tear approximately 30mm distal to the distal edge of the proximal markerband.The balloon material was also torn longitudinally beginning approximately 12mm proximal to the proximal edge of the proximal markerband and extending across the entire length of the balloon material.A microscopic examination on the balloon identified no issues with the balloon material that could have contributed to the complaint incident.None of the balloon material was noted to have detached.A visual and microscopic examination identified no damage or any issues with the tip or markerbands of the device.A visual and tactile examination found no kinks or damage along the shaft of the device.No other issues were identified during the product analysis.As part of the investigation an electronic device history record relating to the batch was performed and no issues were noted with the batch that may have led to the complaint.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
Same case as mdr id 2134265-2017-08300.It was further reported that the target lesion was 90% stenosed, severely tortuous and moderately calcified.
|
| |
|