| Model Number |
VLOCA204L |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, on (b)(6) 2017, intra-operatively, the device needle came out of the endostitch while oversewing the staple line.The needle fell out of endostitch and was retrieved with the grasper.Another device was used to complete the procedure.No patient injury has been noted.The devices are expected to be returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Needle had marks on the cones.Records from each manufacturing lot are thoroughly quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of marks on the cone of the needle may occur when pressure is applied on the toggle lever of the suturing device prior to closing the handles and prematurely attempting to toggle.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|