| Model Number |
DDMC3D1 |
| Medical Device Problem Code |
Invalid Sensing (2293)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/31/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patients right ventricular (rv) lead exhibited a high threshold reading and t-wave oversensing (twos) which was not withheld by the implantable cardioverter defibrillator (icd) t-wave oversensing (twos) discriminator.It was noted the current ekg showed the t-waves were larger than the r-waves.Follow-up was completed with the clinic and the healthcare professional (hcp) verified that reprogramming had been completed.The lead and device remain in use.No patient complications have been reported as a result of this event.
|
| |
|