| Model Number |
ZM-540PA |
| Medical Device Problem Codes |
Electrostatic Discharge (2149); Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The biomedical engineer reported that the nurse was shocked when trying to place batteries into the transmitter.The biomedical engineer then tested the device and the batteries overheated.The batteries were thrown away.There were no signs of physical damage or fluid intrusion.There was no report of patient injury and the nurse was not seriously injured.The biomedical engineer was provided with an exchanged device.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer reported that the nurse was shocked when trying to place batteries into the transmitter.The biomedical engineer then tested the device and the batteries overheated.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer reported that the nurse was shocked when trying to place batteries in the transmitter.The biomedical engineer then tested the device and the batteries overheated.The batteries were thrown away.There were no signs of physical damage or fluid intrusion.There was no report of patient injury and the nurse was not seriously injured.The biomedical engineer was provided with an exchanged device.The investigation has determined the complaint cannot be duplicated.
|
| |
|