| Catalog Number |
7205786 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the device does not work at all times when the handle is pressed.Sometimes though, it will stay permanently on even when switched to safe mode.No injuries or complications were reported as a result.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received and sent to the original equipment manufacturer (oem) for evaluation.There was a relationship found between the returned device and the reported incident.Evaluation of the device by the oem found the unit did not run.When disassembled it was found that the motor had failed and the lohet was cracked.The complaint was confirmed and per the oem the root cause was determined to be wear from use.
|
| |
|