Model Number SC-2316-50 |
Device Problems
Fracture (1260); High impedance (1291); Failure to Align (2522)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/14/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model #: sc-2316-50 serial #: (b)(4) description: infinion 1x16 perc lead kit-50 cm model #: sc-3400-30 serial #: (b)(4) description: infinion splitter 2x8 kit (30 cm) model #: sc-1132 serial #: (b)(4) description: precision spectra implantable pulse generator.
|
|
Event Description
|
A report was received that the patient had high impedances.During revision procedure, the physician had difficulty aligning the lead with the ipg and one of the lead was stuck in the ipg.Lead fracture was suspected.The patient underwent a revision procedure wherein the ipg, leads, and splitters were replaced.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
Additional information was received that the patient was no longer felt stimulation or relief as a result of high impedances.The returned ipg ((b)(4)), clik anchor ((b)(4)), and lead splitter (sc-3400-30/(b)(4)) were analyzed and no anomalies were found.Sc-2316-50 (sn:(b)(4)) device evaluation indicated that the complaint has been confirmed.Visual inspection found lead body was bent/kinked/damaged at the clik anchor site, 1 cm from the set screw mark.The fracture location is 24 cm from the distal end.X-ray inspection confirmed all cables were fractured.There are no exposed cables at the clik site fracture.The fractured cables resulted in the reported high impedances.Sc-2316-50 (sn:(b)(4)) device evaluation indicated that the complaint has been confirmed.Visual inspection found lead body was bent/kinked/damaged at the clik anchor site, 1 cm from the set screw mark.The fracture location is 24 cm from the distal end.X-ray inspection confirmed all cables were fractured.There are no exposed cables at the clik site fracture.Additionally, visual inspection revealed that the lead body has a pulled and exposed cable at the midsection.The fractured cables resulted in the reported high impedances.
|
|
Event Description
|
A report was received that the patient had high impedances.During revision procedure, the physician had difficulty aligning the lead with the ipg and one of the lead was stuck in the ipg.Lead fracture was suspected.The patient underwent a revision procedure wherein the ipg, leads, and splitters were replaced.The patient was doing well postoperatively.
|
|
Search Alerts/Recalls
|